聯(lián)苯
-
乙腈中氟蟲腈砜
BW-NCL-HB-00191 | 100μg/mL
-
乙腈中氧樂果
BW-NCL-HB-00120 | 100μg/ml
-
標(biāo)準(zhǔn)物質(zhì)/乙腈中吡蚜酮
BePure-25305XA-1.2mL | 100μg/mL
-
標(biāo)準(zhǔn)物質(zhì)/乙腈中烯酰嗎啉
BePure-22145XA-1.2mL | 100μg/mL
-
標(biāo)準(zhǔn)物質(zhì)/乙腈中甲霜靈
BePure-23919XA-1.2mL | 100μg/mL
Reference Material
This certificate is designed in accordance with ISO 17034 and ISO Guide 31. This reference material (RM) was designed,produced and verified in accordance with ISO/IEC 17025, ISO 17034 and a registered quality management system ISO9001.
Product Name
Biphenyl
Product Code
BW901063
Lot Number
25041519
CAS No.
92-52-4
Mol. Formula
C12H10
Mol. Weight
154.21
Certified on
21 Apr. 2025
Expiry Date
20 Apr. 2030
Storage Temp
20°C ± 3°C
Specification
100 mg
CERTIFIED
Purity
CERTIFIED
Expanded Uncertainty (U)
Approved by
99.9%
1.0%
Uncertainty
The certified value(s) and uncertainty(ies) are determined in accordance with ISO 17034 with an 95% confidence level (k=2). Uncertainty is based on the Total Combined Uncertainty, including uncertainties of characterisation, homogeneity and stability testing. Stability values are based on real evidence opposed to simulation.
The producer certifies that this reference material meets the specification stated in this certificate until the expiry date, provided it is stored unopened at the recommended temperature herein. Product warranties for this reference material are set out in the terms and conditions of purchase.
Intended Use
This RM is intended for use in a laboratory as a calibration and quality control standard or in method development for analytical techniques.
Safety
Proper precautions should be observed while handling. See Safety Data Sheet.
Traceability
The balances used for gravimetric measurements are calibrated with weights traceable to the national standards (DKD). The calibration of the balances is verified daily internally and annually by an external7 accredited calibration service. Chromatographic methods are traceable to the International System of Units (SI).
Homogeneity
Random replicate samples of the final packaged RM have been analysed to prove homogeneity compliant with ISO 17034.
Storage
The RM should be stored in the original sealed container at the indicated temperature.
Instructions for use
It is recommended to use 1 mg as the minimum sample size and if less material is used, to increase the certified uncertainty by a factor of two for half sample and four for a quarter of sample. If storage after opening is necessary, the RM should be tightly closed and kept from light and moisture. If the RM was in a sealed ampoule, it should be transferred to a vial with minimum head space.
您正在瀏覽的產(chǎn)品:聯(lián)苯
手機(jī)版:聯(lián)苯
本公司銷售的所有產(chǎn)品僅供實(shí)驗(yàn)科研使用,不用于人體及臨床診斷。
2. 熱脫附:脫附溫度300℃,載氣流量40 mL/min;
3. 儀器分析:色譜柱為HP-5MS,程序升溫(40℃起始,5℃/min升至200℃);質(zhì)譜采用SIM模式監(jiān)測(cè)聯(lián)苯特征離子(m/z 154、152)。
2. 聯(lián)苯標(biāo)準(zhǔn)曲線需覆蓋樣品濃度范圍,相關(guān)系數(shù)(R2)≥0.995;
3. 若樣品中聯(lián)苯濃度超出曲線范圍,需稀釋后重新分析。